Blocarea cârligului de extragere în timpul îndepărtării dificile a unui dipozitiv intrauterin Mirena: prezentare de caz

Entrapment of the intrauterine device removing hook during difficult Mirena removal: a case report

Data publicării: 16 Octombrie 2025
Data primire articol: 12 August 2025
Data acceptare articol: 20 August 2025
Editorial Group: MEDICHUB MEDIA
10.26416/Gine.49.3.2025.11102
Descarcă pdf

Abstract

Mirena (levonorgestrel-releasing intrauterine system) is a widely used long-acting reversible contraceptive known for its efficacy and safety in both contraception and ma­nage­ment of heavy menstrual bleeding. While routine re­mo­vals are typically straightforward, various anatomical and technical challenges can complicate the procedure. We present a rare case of mechanical entrapment of an in­tra­ute­rine device (IUD) removing hook during attempted re­trie­val of a Mirena device in a 45-year-old multiparous wo­man. The patient presented for elective Mirena removal after six years of use. On examination, the IUD threads were not visible, and transvaginal ultrasound confirmed that the device was correctly positioned within the uterine ca­vi­ty. During attempted removal using a standard IUD hook, resistance was encountered, and the instrument be­came entrapped within the cervical canal. After the initial at­tempts at gentle manipulation failed, the procedure was paused to reassess the situation. Transvaginal ultrasound ruled out perforation, but suggested cervical stenosis as a likely contributing factor. To facilitate cervical dilation, 200 mcg of misoprostol were administered vaginally. Four hours later, with the cervix softened and partially dilated, the hook was successfully removed under local anesthesia. Sub­sequently, the Mirena was retrieved using hysteroscopic gras­pers under ultrasound guidance without complications. This case highlights the importance of a calm, stepwise ap­proach when managing difficult IUD removals. It also em­pha­sizes the utility of misoprostol for cervical priming and the role of ultrasound and hysteroscopy in safely na­vi­gating complex cases. Instrument entrapment is a rare but important complication that clinicians should be aware of, especially in patients with suspected cervical stenosis or prior uterine surgery. Recognizing early signs of resistance and avoiding forceful extraction are crucial to prevent iatrogenic trauma.



Keywords
Mirenaintrauterine device (IUD)IUD removalcervical stenosismisoprostolultrasound guidanceinstrument entrapmenthysteroscopy

Rezumat

Mirena (sistem intrauterin cu eliberare de levonorgestrel) este un contraceptiv reversibil de lungă durată, utilizat pe sca­ră largă, cunoscut pentru eficacitatea și siguranța sa atât în con­tra­cep­ție, cât și în managementul menstruațiilor abun­den­te. Deși îndepărtarea de rutină este, de regulă, simplă, anu­mi­te particularități anatomice și dificultăți tehnice pot com­pli­ca procedura. Prezentăm un caz rar de blocaj mecanic al cârligului de extracție al dispozitivului intrauterin (IUD) în tim­pul unei încercări de îndepărtare a unui dispozitiv Mirena la o pacientă multipară în vârstă de 45 de ani. Aceasta s-a pre­zen­tat pentru extragerea electivă a dispozitivului Mirena după șase ani de utilizare. La examenul clinic, firele IUD nu erau vi­zi­bi­le, iar ecografia transvaginală a confirmat poziționarea co­rec­tă a dispozitivului în cavitatea uterină. În timpul încercării de extragere cu un cârlig standard pentru IUD, medicii au în­tâm­pi­nat rezistență, iar instrumentul a rămas blocat în ca­na­lul cervical. După încercările iniţiale de manipulare blândă, pro­ce­du­ra a fost întreruptă pentru reevaluare. Ecografia trans­va­gi­na­lă a exclus perforația, sugerând stenoza cervicală ca posibil fac­tor determinant. Pentru a facilita dilatarea colului, s-au ad­mi­nis­trat intravaginal 200 mcg de misoprostol. După patru ore, cu colul uterin înmuiat și parțial dilatat, cârligul a fost extras cu succes, sub anestezie locală. Ulterior, dispozitivul Mirena a fost în­de­păr­tat utilizând pensa histeroscopică, sub ghidaj eco­grafic, fără complicații. Acest caz evidențiază importanța unei abordări eta­pi­za­te în gestionarea îndepărtării dificile a unui dispozitiv in­tra­ute­rin. De asemenea, subliniază utilitatea miso­pros­to­lului pentru pregătirea colului, precum și rolul ecografiei și al his­te­ro­sco­piei în gestionarea situațiilor complexe. Blocarea unor dis­po­zi­tive medicale reprezintă o complicație rară, dar im­por­tan­tă, pe care clinicianul trebuie să o aibă în vedere, în special la pacientele cu stenoză cervi­ca­lă suspectată sau cu antecedente de intervenții uterine. Recunoașterea precoce a semnelor de rezistență și evitarea extragerilor utilizând forța sunt esențiale pentru prevenirea traumatismelor iatrogene.

Cuvinte Cheie
Mirenadispozitiv intrauterin (IUD)îndepăr­tare IUDstenoză cervicalămisoprostolghidaj ecograficblocaj al instrumentelorhisteroscopie

Introduction

Intrauterine devices (IUDs), particularly the levonor­gestrel-releasing intrauterine system (Mirena), are among the most effective forms of reversible contraception, being increasingly utilized for non-contraceptive indications such as menorrhagia(1,2). With over 180 million users worldwide, IUDs offer long-term protection, with high satisfaction rates and low complication profiles. Despite the generally safe and uncomplicated nature of IUD insertions and removals, certain clinical scenarios can pose significant challenges. These include postmenopausal status, nulliparity, previous cervical procedures and uterine anomalies, all of which may contribute to difficulty in locating or extracting the device. The absence of visible IUD threads further complicates the process, often necessitating imaging or more invasive techniques for successful removal. While several strategies exist for managing difficult intrauterine device removals – such as cervical ripening agents, ultrasound guidance and hysteroscopy –, the entrapment of the removal instrument itself is a rare but notable complication. This situation requires careful clinical judgment to avoid iatrogenic injury, while ensuring complete removal of both the device and the entrapped tool.

We present a unique case involving the entrapment of an IUD removing hook during Mirena retrieval. The case illustrates the importance of recognizing early signs of resistance, employing appropriate pharmacologic and procedural interventions, and maintaining a structured, minimally invasive approach to ensure optimal outcomes.

Case report

A 45-year-old multiparous woman presented to the outpatient gynecology clinic for elective removal of a Mirena intrauterine device that had been inserted six years earlier for the management of idiopathic menorrhagia. The patient reported no current symptoms, and she had no history of pelvic inflammatory disease, prior uterine surgery or cervical dysplasia. She had regular menstrual cycles, and she was not using any other form of hormonal contraception. On speculum examination, the cervix appeared normal, but no IUD threads were visible. A bimanual exam revealed a normal-sized anteverted uterus with no tenderness or adnexal masses. Given the absence of visible threads, a transvaginal ultrasound was performed to confirm the presence and position of the Mirena. Ultrasound demonstrated the device located in the mid-cavity of the uterus, with no evidence of displacement or perforation. Under aseptic conditions and with informed consent, an attempt was made to remove the Mirena using a standard IUD removing hook. The hook was introduced through the cervix into the uterine cavity and carefully maneuvered to engage the stem of the IUD. However, upon attempting to withdraw the hook, firm resistance was encountered. Multiple gentle attempts at repositioning and rotating the hook failed to release the device or allow withdrawal of the instrument. Given concerns about possible cervical stenosis or uterine spasm, the procedure was temporarily halted to avoid causing trauma. The patient remained asymptomatic, with no signs of distress or pain beyond mild discomfort. A repeat transvaginal ultrasound was performed to assess the position of both the IUD and the hook. The imaging showed the hook adjacent to the IUD stem, without evidence of uterine perforation or displacement of the device. After discussing the situation with the patient, it was decided to administer 200 mcg of vaginal misoprostol to promote cervical softening and dilation. The patient was advised to return four hours later for the continuation of the procedure. Upon return, the cervix was noted to be softer and slightly more dilated on physical examination. Under local paracervical block, the hook was gently rotated and withdrawn without resistance. Subsequently, a hysteroscope was introduced into the uterine cavity under ultrasound guidance. Using grasping forceps, the Mirena was successfully removed, without complications. The patient tolerated the procedure well, and she was discharged home with instructions to monitor for any signs of infection or abnormal bleeding. The follow-up one week later confirmed no adverse events.

Discussion

Difficult IUD removals are relatively common in clinical practice, particularly in postmenopausal women, nulliparas, or those with prior cervical surgeries such as conization or loop electrosurgical excision procedure (LEEP)(3). These populations often exhibit cervical stenosis or reduced cervical elasticity, making access to the uterine cavity challenging. Additionally, factors such as retroverted uteri, embedded IUD arms, or fibroid distortion can further complicate the procedure. Instrument entrapment during IUD removal is an infrequent but important complication that has received limited attention in the literature(4). To our knowledge, only a handful of cases have been reported where the removal hook became lodged during the procedure. Possible mechanisms include improper angulation of the hook, cervical narrowing creating a mechanical barrier, or unintentional engagement of the hook with endometrial tissue rather than the IUD itself.

In this case, several key principles guided the successful management of the complication:

1. Recognition of resistance and avoidance of force

Early recognition of resistance is critical. Continuing to apply force risks cervical laceration, endometrial damage, or even uterine perforation. Halting the procedure allowed time for reassessment and planning without compromising patient safety.

2. Use of imaging for accurate assessment

Transvaginal ultrasound played a pivotal role in confirming the location of both the IUD and the hook, ruling out perforation, and guiding subsequent steps. Real-time imaging helps clinicians visualize the anatomy and avoid blind manipulations.

3. Cervical priming with misoprostol

Misoprostol, a synthetic prostaglandin E1 analog, is widely used for cervical ripening in various gynecologic settings, including IUD removal. Its effectiveness in softening and dilating the cervix has been well documented, especially in postmenopausal women or in those with stenotic cervices. In this case, the administration of 200 mcg vaginally led to sufficient cervical changes that facilitated the safe removal of the hook and IUD.

4. Stepwise escalation of techniques

Rather than resorting immediately to operative hysteroscopy or general anesthesia, a gradual escalation from basic to advanced techniques minimized procedural burden and preserved fertility potential(4). The use of local anesthesia and targeted ultrasound-guided intervention avoided unnecessary exposure to systemic sedation.

5. The role of hysteroscopy

Hysteroscopy provides the direct visualization of the uterine cavity and allows for precise manipulation of instruments. In cases where standard removal methods fail, hysteroscopy is considered the gold standard for safe and effective IUD retrieval(5). Here, it enabled the secure engagement and removal of the Mirena once the hook was dislodged.

Clinical implications and recommendations

This case underscores several important clinical takeaways:

Clinicians should maintain a high index of suspicion for cervical stenosis in patients with absent IUD threads or difficulty accessing the uterine cavity.

Early use of cervical ripening agents like misoprostol can significantly improve the procedural success.

Imaging modalities, particularly transvaginal ultrasound, are essential tools in assessing IUD position and guiding interventions.

Instrument entrapment, although rare, should be recognized as a distinct complication requiring careful handling.

A stepwise approach – starting with conservative measures and escalating only as needed – is sa­fer and more cost-effective than immediate surgical intervention.

Finally, clinician experience and familiarity with advanced removal techniques play a vital role in managing these cases effectively. Training programs should emphasize the recognition and management of rare complications such as instrument entrapment to ensure preparedness in clinical practice.   

 

Data availability. All data supporting the findings are included within the article. No additional source data are required.

Competing interests. The authors declare no competing interests.

Acknowledgments. We thank the patient for her cooperation and the staff of the gynecology unit for their support.

 

Corresponding author: Smit Bharat Solanki E-mail: drsmitbharat@gmail.com

Conflict of interest: none declared.

Financial support: none declared.

This work is permanently accessible online free of charge and published under the CC-BY licence.

Bibliografie


1. Selected Practice Recommendations for Contraceptive Use. 3rd edition. Geneva: World Health Organization; 2016. https://www.who.int/publications/i/ item/9789241565400
2. Faculty of Sexual and Reproductive Healthcare. Intrauterine Contraception Clinical Guidance. FSRH; 2019.
3. Prine L, Shah M. Long-acting reversible contraception: difficult insertions and removals. Am Fam Physician. 2018;98(5):304-309.
4. Brunie A, Aw FNRS, Ndiaye S, et al. Making Removals Part of Informed Choice: A Mixed-Method Study of Client Experiences with Removal of Long-Acting Reversible Contraceptives in Senegal. Glob Health Sci Pract. 2022;10(5):e2200123.
5. Christofield M, Lacoste M. Accessible contraceptive implant removal services: an essential element of quality service delivery and scale-up. Glob Health Sci Pract. 2016;4(3):366-372.

Articole din ediția curentă

EVENTS

September-December 2025 Calendar

The 13th Congress of the Romanian Society of Ultrasound in Obstetrics and Gynecology (SRUOG)...
GYNECOLOGY

Terapia de substituție hormonală la menopauză: review narativ actualizat privind beneficiile, riscurile şi recomandările clinice

Adelina-Georgiana Vârciu
Introducere. Menopauza, determinată de scăderea producției de estrogen, este asociată cu modificări vasomotorii, urogenitale, osoase, cardiovasculare și neurocognitive....
GYNECOLOGY

Impactul chirurgiei endometriomului ovarian asupra fertilității

Irina Burdeniuc, Hristiana Caproș, Ion Bologan, Corina Darii, Ion Burdeniuc
Introducere. Endometriomul ovarian este o manifestare frecventă a endometriozei și o cauză importantă de infertilitate....
Articole din edițiile anterioare

GYNECOLOGY

Terapia de substituție hormonală la menopauză: review narativ actualizat privind beneficiile, riscurile şi recomandările clinice

Adelina-Georgiana Vârciu
Introducere. Menopauza, determinată de scăderea producției de estrogen, este asociată cu modificări vasomotorii, urogenitale, osoase, cardiovasculare și neurocognitive....
GYNECOLOGY

Impactul chirurgiei endometriomului ovarian asupra fertilității

Irina Burdeniuc, Hristiana Caproș, Ion Bologan, Corina Darii, Ion Burdeniuc
Introducere. Endometriomul ovarian este o manifestare frecventă a endometriozei și o cauză importantă de infertilitate....
GYNECOLOGY

Două declanșări, niciun ovocit: un caz rar de sindrom autentic al foliculilor goi rezolvat cu protocolul de declanșare duală

Smit Bharat Solanki
Sindromul foliculilor goi (empty follicle syndrome; EFS) este un fenomen neobișnuit, dar frustrant, observat în cadrul tehnicilor de reproducere asistată (ART), în care nu se obțin ovocite în ciuda un...